Care of Accessories: The hose should last 3-5 years or until an air leak is reported on monitor or hose is damaged. All cuffs should have good sticking Velcro and no holes or gashes in the tubing. Never tape a cuff to keep it on and never tie off the hose to avoid a hole. These cuffs should be replaced as needed.
Cleaning: You can use a lint roller to clean hair off disposable cuffs. You can use hydrogen peroxide to clean blood off disposable cuffs. To clean reusable cuffs and hose use mild detergent/cleaner
Best Practices:
Attachments: : Keep cable and sensors attached to monitor and store in no less than a 6 inch circle, and store next to the monitor.
Keep cable and sensors attached to monitor and store in no less than a 6 inch circle, and store next to the monitor.
Placement: Readings are best taken on a front limb, closest to the heart. Make sure to stay within the index line inside the cuff and follow artery line on the outside of the cuff. For awake cats, the tail can be used to take readings.
Methods: Make sure cuff is snug on the patient, if it slides down or if module is timing out and not giving a reading, go down a cuff size. If cuff pops off patient when inflated, go up a cuff size
Troubleshooting:
If the NIBP is not reading or timing out:
1. Check to be sure the NIBP is turned on. To check this, click on the NIBP tab or in the NIBP portion of the screen and make sure this parameter is on. If the NIBP is turned on and the module is not inflating cuffs, please load factory default settings. Click on System, then select Load Factory Default Settings in the menu. You will see a prompt asking if you are sure say yes. If the NIBP still does not inflate, contact Digicare for service
2. Check your hose and cuffs for holes or gashes. If the Velcro is worn or the cuffs are being taped on, please dispose of those cuffs and replace with new ones.
3. Check to be sure you are in the correct mode. At the top of the screen below the battery gauge, you will see a drop down menu. At the top right corner of the screen below the battery gauge you will see a drop down menu where you can select large, small or rodent (LW8 only). Large mode is used for patients over cuff size 3, small is for patients cuff size 3 and under and rodent for exotics.
4. If the module inflates but times out, change your cuff one size smaller.
Last Modified: 04/23/2019