MRI Setup
Customer Service 561-689-0408

MRI Setup

Please read manual thoroughly prior to installing your LifeWindow monitor as proper installation is crucial for best results.

 

**The MRI monitor must be placed in the control room outside of the MRI room and the screen should be facing the window into the MRI room so that the alarms can be seen.

**The ECG leads MUST go through the patch panel separate from other accessories and directly to the MRI unit. This is important to prevent interference.

**Sp02, CO2 and NIBP accessories MUST go through the waveguide and directly to the MRI unit.

**It is recommended to consult with a technician for installation.