Quick Reference Guide
To turn on, double click black switch in the back. Turning off monitor, use blue system tab then select shut down monitor from the menu. When the silence alarm button is pressed one time, the alarm will be silenced for 2 minutes. If you wish to silence longer, press and hold the silence alarm button until “audible alarms” stops flashing in the middle of the screen. LifeWindow monitors are networkable meaning you can save your data directly to a folder on your hospital computer and easily upload it to your practice management software. Please see networking section in the knowledge database for instructions on this.
ECG:
• We recommend using gel to attach the leads for the best contact as alcohol dries out
• Using large and small mode is very important for ECG readings. This setting is in the top right corner of the screen just below the battery info. For patients<30# use in small mode, for patients >30# use in large mode. You may need to toggle between leads for patients in the 20-30# range
• Lead II is preferred but if double counting occurs, using the ECG button on the keypad you can toggle in between leads I, II and III. For Lite monitors (LW8), you can also use rodent mode for the smaller patients under 30#
• Make sure Mains Notch and Digital are always on. This will filter out interference and prevent double counting
• Do not allow the leads to touch each other or any metal surface. If patient is lateral, try rolling a towel and placing between legs to try and create distance between leads.
• Be sure to use the tension clip to prevent the cable from pulling and always unclip, never pull them off a patient
• Replace leads and cable if fraying or cuts are seen
• Alarm limits can be changed by clicking the blue ECG tab
Sp02
• For monitors with Masimo, you can change sensitivity if you have trouble getting readings on small patients. To do this, click on Sp02, Sp02 again then in the left column click on the dropdown menu to change sensitivity to maximum. You can also change the sensitivity to maximum for low perfused or critical patients.
• Readings can be obtained from ear, tongue, vulva, prepuce or between toes. We do not recommend having the sensor on the tongue during dentals as they are sensitive and will be damaged by the water. We have a reflectance sensor that is used on the base of the tail and is great for use in dentals, please see your accessory care sheet for more info on this.
• If Sp02 is reading low on a patient you don’t think it should be, try changing mode to normal
• Move sensor off tongue every 20 minutes to prevent pressure necrosis
• Make sure sensor is parallel on the patient and use the correct clip size. You should have 2 different sizes, the sensor should not be clamped down or open wide. For small patients, you can also fold the tongue or use a gauze square between the tongue and sensor for better readings.
• Clean sensor with dilute alcohol on a gauze square. Replace sensor and cable if fraying or cuts are seen
• Alarm limits can be changed by clicking the blue Sp02 tab
NIBP:
• Readings are best taken on a front limb, closest to the heart
• Make sure to stay within the index line inside the cuff and follow artery line on the outside of the cuff
• Large and small mode is extremely important for accurate readings. For patients using cuff sizes under 3, use in small mode, for patient using cuffs over 3, use in large mode
• Make sure cuff is snug on the patient, if it slides down or if module is timing out and not giving a reading, go down a cuff size. If cuff pops off patient when inflated, go up a cuff size
• If Velcro is worn, please replace cuff, never tape a cuff
• Alarm limits can be changed by clicking the blue NIBP tab
CO2:
• Sidestream: We have two different types of sample lines available. Low humidity lines are recommended for dentals and high humidity lines are recommended for routine surgeries. There are two sizes, pediatric (purple) is for trach tubes under 4 and adult (clear) is for trach tubes over 4
• Change lines in between patients, hang them with the mouth piece down to allow them time to dry out
• Perform a zero calibration every time a sample new line is placed, when switching between sizes, if you receive the no breaths message or readings seem off. This must be done off the patient and input exposed to 0 mmHg CO2 concentration. To do this, click on the CO2, then Zero. The monitor will ask if it is exposed to 0 mmHg, click yes. You will see “zero in progress” on the screen then when the zero is finished, the numbers will flash and you are ready to use. If you receive a “zero error” message, repeat the process or try a new sample line
• If you receive an “occlusion” message, replace sample line and do zero calibration. Sample lines and airway adapters are disposable items and need replaced approximately every 2-3 months. If a sample line has a cut or break in it, throw it away and replace with a new one. We recommend having 1-2 extra for when this happens.
Mainstream: We have two different size airway adapters, pediatric (purple) for trach tubes under 4, adult (clear) for trach tubes over 4
• Perform a zero calibration every time a new airway adapter is placed, when switching between sizes, if you receive the no breaths message or readings seem off. This must be done off the patient and input exposed to 0 mmHg CO2 concentration. To do this, click on the CO2, then Zero. The monitor will ask if it is exposed to 0 mmHg, click yes. You will see “zero in progress” on the screen then when the zero is finished, the numbers will flash and you are ready to use. If you receive a “zero error” message, repeat the process or try a new airway adapter.
• If you receive a “check adapter” message, replace airway adapter and do a zero calibration. Airway adapters are disposable items and need replaced approximately every 2-3 months. If an airway adapter is cracked or has been damaged, please replace. We recommend having 1-2 extra for when this happens.
• Alarm limits can be changed by clicking the blue CO2 tab
Printing and saving patient data:
• When the monitor is turned on, it is automatically recording patient data. When you admit a patient, the data transfers the data to the patient’s buffer. The patient can be admitted any time after starting the monitor and data from the buffer can be saved or discarded
• The LifeWindow monitor can print a 30 second waveform Waveform print options available by printing from a strip chart (optional), saving to a thumb drive or connecting an external printer.
• The LifeWindow monitor can save tabular reports to an Excel file and are recorded every minute. The report can be reconfigured at the time of printing or after an Excel has been generated and readings can be taken between 1 and 99 minutes. A patient must be admitted in order to save the Excel report
• To save the patients data, click on System then Save Patient Vital Signs Report and choose where to save to. You can save locally (d:/Patient Data) to the monitor or to a thumb drive (E). We always recommend saving this way before discharging to avoid accidental loss of data.
Battery:
• To preserve battery life, once a month turn monitor on and let the battery drain until it shuts off. Plug back in to recharge.
Storing accessories:
• We recommend keeping accessories attached to the monitor at all times. It is best to loosely wrap the ends and store next to the monitor in a basket or drawer. Never wrap or wind them tightly around an object.
General care and cleaning:
• To clean the touch screen, use a soft cloth with a small amount of a mild cleaner to clean off any debris and dry well. Take care not to let any moisture seep into the corners and edges
• Accessories can be cleaned with dilute alcohol on cotton or gauze. Take care when cleaning Sp02 and CO2 sensors to prevent damage