Admitting a Patient:
You will need to admit a patient in order to save data. Admitting a patient can be done on the iPad or the monitor. To admit from the iPad, click Home, Patient Admission, enter patient information and Admit. You can then add a photo from the gallery or take a new one. You are now ready for monitoring. You can add notes to your patient’s records by clicking in the notes box at the bottom of the iPad.
Monitoring:
You can change parameter settings and alarm limits by touching on that parameter. The monitor includes five different patient types (Canine, Feline, Equine, Rodent, Exotic). Canine, Feline, and Exotic also include an option for Large/Small. Changing the patient type affects the NIBP and ECG functions of the monitor. It is important to select the correct patient type when monitoring patients to ensure both patient safety and measurement accuracy. It is important to use small mode for patients under 30 lbs or cuff size 3 and under and large mode for patients over 30 lbs or cuff size over 3.
Accessories:
Keep cables and leads attached to monitor and store in no less than a 6 inch circle, and store next to monitor. The only accessory that should be removed daily is the CO2 sample line if your LifeWindow One has this parameter. Clean leads, cable and clip with a mild cleaner/disinfectant on a damp cloth or gauze and dry well. Accessories will need to be replaced over time as they do wear out. CO2 sample lines (sidestream) airway adapters (mainstream) will need to be replaced the most frequently (every 2-3 months or when occlusion is seen). Sp02, ECG and NIBP will need to be replaced every 1-5 years depending on use and care.
Saving ECG Waveforms:
Touch on the ECG parameter and select Capture under ECG Recording. You can now save 30 seconds and up to 2 hours of the waveform by using the slider bar or the -/+ sign and choosing Save. The iPad will generate the file and the record can now be viewed, printed or emailed under the ECG Records tab on the Home screen of the iPad. Select your patient and click on View. From here you can Extract (Email), Print or Delete the file. If you would like to email the file, you will need to click on Extract, click on Mail. You can also air drop to another iPad to save to another app.
Saving Tabular Records:
When you are ready to discharge your patient, click on Home then Discharge and the file will be saved to your iPad. To view, print or email, click on Patient Records from the Home screen, select your file and click on View then choose Extract or Print.
Settings Screen:
[1] Monitoring – Returns to the main monitoring screen.
[2] Patient – Selection for the patient type.
[3] Trends – Displays the tabular and graphical trends information.
[4] General – System settings for Display, Date and Time, etc.
[5] Alarms – Selection of the alarm limits for each numeric.
[6] Sounds – Selection for the volume of the pulse beep and alarm beep.
[7] Parameter – Configuration for each individual parameter numeric.
[8] Waveform – Configuration of each available waveform.
[9] CO2 Settings – CO2 Module settings.
[10] Temperature Settings – Temperature Module settings.
[11] SpO2 Settings – SpO2 Module settings.
[12] NIBP Settings – NIBP Module settings.
[13] About – Monitor information including S/N, software version, release date, etc.
Last Modified: 04/22/2019