IP41x FAQ's and Best Practices
Customer Service 561-689-0408

IP41x FAQ's and Best Practices

IP41x FAQ’s

How do I change the drip count for my IV line?

Click Home, System Set, Maintenance, Local Set, Enter password 1666 and press confirm, Consumable Setup, Calibrate, Edit size and press confirm. You will see the new drip rate at the top of your home screen.

How do I make the Anti free flow clamp stay open?

Click on Home, then System Set. Scroll down to uncheck linkage mode.

Why am I getting an occlusion alarm?

The occlusion alarm can occur when a high-viscosity liquid is infused at a high speed through a thin needle or catheter. You can increase the occlusion by clicking P at the top of the screen or by decreasing the infusion speed.

How can I turn off the power supply reminder alarm?

Click on Home, System Set, No AC power supply alarm, enter password 1234 and press confirm. Uncheck the no AC power supply alarm box.

What is purge mode used for?

Purge is to be used before infusion to purge air bubbles from the line before connecting to the patient.

How do I adjust the volume?

Click on Home, System Set, Volume Setting, enter password 1234 and press confirm. Use slide bar to adjust volume and press confirm.

Best Practices:

The roller and pinch clamps must remain above the pump at all times.

Select the correct brand and drop rate of your IV set.

Keep the pump plugged in at all times.

Ensure pump is on a stable counter or IV pole.

Standby mode can used if a patient needs to go for a walk or a treatment or is being discharged but returning at a later time. To enter standby mode, simply click on the blue power button and choose standby on the touch screen. This will stop all audible alarms for up to 24 hours and allow the technician to resume therapy without interruption. When the patient is reconnected to the pump, the user then just needs to press cancel to resume fluid therapy.

It is important to set the VTBI to the amount of fluids you want to be administered over a set time, not the amount of fluids in your bag.

Double check your decimal point before starting infusion!

Do not use this pump for blood transfusions. The SR31x is approved for transfusions if needed.

Monitor your patient consistently to ensure fluids being administered is the correct amount that you set the pump to deliver.

Always check on your patient during infusion and measure the volume that has been administered to ensure accuracy.

 

 

 