ACCESSORY CARE FOR LIFEWINDOW 9X AND 8X

    **All accessories can be cleaned with dilute alcohol or a mild cleaner

  

 ECG (Keep attached to monitor and hang leads and cable) 

LEADS: For these monitors you can either have the 3 lead alligator clips (EC003) or the 3 lead snap in smooth clips (EC014). ECG leads should be replaced as needed or approximately every 2 years to maintain good contact.  CABLE: This should last for a long time unless it is damaged or chewed through. Every 3-5 years. (EC001) 

SPO2 (Keep attached to monitor and hang sensors and cable) 

  • SENSOR: The technology of this SpO2 is Masimo.  Please be sure when using lingual area (PO756 or PO758) to relocate the sensors position every 30 minutes to avoid pressure necrosis. This sensor will usually last about 8-12 months depending on use.  You may see a sensor error on the screen of the monitor when this sensor needs to be changed.     
  • Optional Sensors:  Tip Clip PO756 Transflectance: PO757
  • Cable: Replace as needed. Every 3-5 years (PO759) 

SIDESTREAM CO2 (This is the only accessory you should remove daily from the monitor) 

  • The CO2 technology is Respironics. There are 4 different sample lines that can be used. All sample lines are good for 72hrs of use or 2-3 months. We recommend when you open a new sample line to use tape and put the date on the sample line.
  • All sample lines should be rotated between surgeries to give the line time to dry. Hang the lines with mouth piece down and filter to the ceiling to allow all the moisture to drain down to the mouth piece. - Whenever you open a new sample line or if you are measuring the patient and the readings seem a little off you must do a zero calibration. Make sure the sample line is off the patient and you want to hit the zero calibration button located in the CO2 parameter of you monitor. Once it reports it was successful the line is ready to be used.
  • High humidity (CAP120) for trach tubes over 4. Recommended for general surgery 
  • High humidity, low dead space (CAP123) for trach tubes under 4. Recommended for general surgery 
  • Low humidity (CAP121) for trach tubes over 4. Recommended for dentals 
  • Low humidity, low dead space (CAP122) for trach tubes under 4. Recommended for dentals 
  • Luer lock/Low Humidity Equine (CAP124) Trach tube above 4, Recommended for dental cleanings or oral surgery 
  • Luer lock/High Humidity Equine (CAP125) Trach tube above 4. Recommended for general surgery 

MAINSTREAM CO2: (Keep mainstream sensor attached to monitor and remove adapters only)

Though considered disposable (CAP112&113), in the veterinary field these are reused and recommended to be changed every

4-6 months. Please note if the adapter becomes broken or cracked it must be changed immediately and it is always best to

keep extra on hand in your clinic.

- Airway Adapter- Disposable Regular Dead Space: Trach tube above 4, all surgeries and dentals(CAP112)

- Airway Adapter- Disposable Low Dead Space: Trach Tube under 4, all surgeries and dentals (CAP113)

- Airway Adapter – Low Dead Space Reusable: Trach tube under 4, all surgeries and dentals (CAP111) *Autoclavable*

- Airway Adapter – Regular Dead Space Reusable: Trach tube above 4, all surgeries and dentals (CAP112) *Autoclavable*

- Please perform zero-calibration when moving from regular dead space to low dead space adapters and vice versa. *Please

  • find directions for zero-calibration above*

 

NIBP (Keep attached to monitor and wrap loosely next to monitor or in a basket or drawer)  

HOSE: (BP013) The hose should last 3-5 years or until an air leak is reported on monitor or hose is damaged. 

CUFFS: (BP014-BP024) All cuffs should have good sticking Velcro and no holes or gashes in the tubing. Never tape a cuff to keep it on and never tie off the hose to avoid a hole. These cuffs should be replaced as needed.

 

 TEMP (Keep attached to monitor and hang)  

-REUSABLE TEMP PROBE-(TP004) This probe has a sensor at the tip of the probe. Please be careful when placing this sensor to not hit hard feces or a wall in the rectum. This sensor usually last about 8-12 months depending on use. If tip of sensor is cracked or broken, please replace. Plastic disposable probe covers are available in boxes of 100, (PO732)

Masimo Transflectance Sensor

 

The transflectance sensor is a great option for use in dental procedures or oral surgery as it attaches to the base of the tail just above the rectum using Vetrap. It is important that the emitter and receiver touch directly onto skin so you might have to do a light clip of hair around the ventral tail. Use vet wrap to place the sensor on the ventral base of the tail either transversely on a wide based tail or longitudinally if a narrow based tail so that it can hit a coccygeal bone and bounce back.  

The transflectance sensor requires more technique to use than the lingual sensor. 

Tips for using the transflectance sensor are: 

If you wrap this sensor too tightly, you may impede perfusion to the sensor site. Wrap this sensor so that it is  “snug”. If the sensor is wrapped too loosely, it may move and display erroneous readings. 

Optional Site - TFI sensor on the metatarsal*. There is a spot immediately behind the large pad that has a slight indentation. Remove the hair and wrap the ankle securely, without too much pressure, and wait 10 to 15 seconds to see if you get a good pulse signal every time the heart beats. Utilize the PI and SIQ bar to help optimize sensor placement. If SIQ and PI are RED reposition sensor or try another site. If SIQ and PI are GREEN secure sensor in place with wrap. 

PVI, SpO2, HR, and PI monitoring in Small Dogs (less than 60-70 lbs.) using TCI on the Tongue 

If you have a small size dogs (less than 60-70 lbs.), then double folding their tongue to itself so that you can increase the thickness of the tongue tissue. After such folding, then place the lingual probe over the thickening tongue. It is essential that the emitter and receiver are parallel. You should be able to get better PI and roughly after 3-4 minutes later you can get readings from PVI. 

PVI, SpO2, HR, and PI monitoring in Large Dogs using TCI on the Tongue 

When monitoring SpO2, HR, and PI only in Large Dogs (greater than 70 lbs.) place the Lingual TCI clip sensor on the tongue. Make sure the tip of tongue is touching the inner base of sensor and ensure the emitter and receiver are parallel.   

PVI, SpO2, HR, and PI monitoring in Large Dogs using TCI on the Tongue 

 When monitoring SpO2, HR, and PI only in Large Dogs (greater than 70 lbs.) place the Lingual TCI clip sensor on the tongue. Make sure the tip of the tongue is touching the inner base of the sensor and insure that the emitter and receiver are parallel. 

Last Modified: 04/12/2019