1. Enable and Select IV set brand.
1.1. The IP41x has several IV line brands stored in the library. To add a new IV line you need
to edit and calibrate one of the brands that are not being used.
1.2. Press “Home” key to enter the Home screen.
1.3. Press “System Set” and press “Collection Set”.
1.4. Press “Brand Collection”. Check the box to enable any unused brand in the library.
1.5. Press “Home” key twice to enter the Home screen.
1.6. Press “Infusion Set”, press “Brand” and select the Brand you want to edit or calibrate.
2. Edit and Calibrate IV Set brand
2.1. Press “Home” key twice to enter the Home screen.
2.2. Then press “System Set” and press “Maintenance”.
Press “Local Set”, enter password “1666”, click “Confirm”
2.4. Press “Consumable setup”.
2.5. Press “Calibrate”.
2.6. Press “Edit”.
2.7. Press “Producer”. Use the arrow back key to erase the current Brand and type the new IV set
Brand and press “confirm”. Repeat same procedure to edit the new IV Model and Size. Size is
the milliliters (ml) per drop specification.
Press the arrow back key to return to the “Calibrate” page. Install the IV Set in the pump.
Use “Start Purge” to get the disposable prepared for the calibration, remove all the bubbles,
and prepare a calibrated cylinder at the end of the IV set to measure the infused volume
during the calibration infusion. Press “Stop Purge” to stop the purge.
Make sure there is enough liquid in the IV bag for the calibrating volume of 15.70 ml or other
automatically set volume for the calibration test.
2.9. Click “Start/Stop Cal.” to start. The screen shows “Calibrating time” as below. The calibration
infusion takes 10 minutes.
2.10.Ten minutes later, the calibration finishes automatically. Measure the volume of the measuring
cup and enter the measured volume replacing the 15.70 ml or other volume originally in the
area showing below. At the end of the calibration press confirm.
The calibration is done. Press Home to return to the main page.
A digital scale can also be used to precisely measure the infused volume by the calibration
infusion.
Measure first the weight of the empty container: W1
Measure the weight of the container with the infused water by the calibration infusion: W2
Use the following formula to calculate the infused volume: V = (W2-W1)/0.998
Note: 0.998 is the density of the purified water.
Click Return to return to calibration screen.