SpO2 - Replace Cable
Customer Service 561-689-0408

SpO2 - Replace Cable

Troubleshooting:

1. Replace cable

https://www.digi-vet.com/Accessories/Monitoring/Pulse-Oximetry/LifeWindow-LW8-Lite/Masimo-SpO2-Patient-Cable.html

 

Care of Accessories: This sensor will usually last about 8-12 months depending on use.  You may see a sensor error on the screen of the monitor when this sensor needs to be changed. Replace the cable as needed about every 2-5 years. Replace sensor and cable if fraying, cuts or visible damage are seen

Cleaning: Clean the sensor and cable with a mild cleaner/disinfectant and dry well. Use care when cleaning the sensor and don’t clean with anything abrasive, a Kim wipe or cotton is okay. Never submerge sensor in fluid or cleaning solution.

Best Practices:

Attachments: Keep cable and sensors attached to monitor and store in no less than a 6 inch circle, and store next to the monitor.

Placement: The Y and tip clip can be used on the tongue, vulva, prepuce, digits or pinnaes. Transflectance sensor can be used on the base of the tail and is attached with Vet wrap where it will be placed length wise so that it can hit the coccygeal (COCK SI GEAL) bone and bounce back. This sensor can also be used on then metatarsal. This is the spot immediately behind the large pad that has a slight indentation. You may need to clip a bit of hair for this and attach using Vet wrap.   

Methods:

Once you place your sensor wait about 10-15 seconds for a reading. Also, use your PI and SIQ bar to help optimize sensor placement. If SIQ is red and PI is low, reposition the sensor or try another site. If SIQ is in green and the PI is high your reading is accurate.

When using tip-clip or Y sensor it is highly recommended to move sensor every 20 mins to avoid pressure necrosis.

Make sure the sensor is parallel on the patient not open wide or clamped down. Gauze can be placed in between the tongue and sensor for patients with small or thin tongues.

 

  • You can change sensitivity if you have trouble getting readings on small patients. To do this, click on Sp02, Sp02 again then in the left column click on the dropdown menu to change sensitivity to maximum. You can change the sensitivity to maximum for low perfused or critical patients.
  • Your SIQ should always be in green
  • Your PI – Perfusion Index tells you how much blood flow is going to that site. There is no perfect number for PI. The higher the number the better. Your smaller patients will tend to have a lower PI than your larger patients. If your PI continues to be low change sensor location.
  • Do not use the sensor on the tongue during dentals as they are sensitive and will be damaged by the water.
  • Patient with thin tongues and ears you can fold them or use gauze in half or quarters
  • Make sure monitor does not report replace cable or replace sensor

Last Modified: 04/12/2019