Troubleshooting:
1. Replace airway adapter and perform zero calibration.
Care of Accessories: On average, the airway adapters last around 4-6 months, less if you are a busy practice. If an airway adapter is cracked or has been damaged, please replace.
Cleaning: The airway adapters can be removed from the mainstream sensor and placed in the chlorhex solution to disinfect and clean, make sure to fully dry.
Best Practices:
Attachments: Keep mainstream sensor attached to monitor. Remove airway adapters and clean in between patients.
Placement: The mainstream sensor will be placed next to the patient on your surgery table, please use care when handling this and don’t allow it to fall or get wet. You will use the pediatric (purple) for trach tubes under 4, and the adult (clear) for trach tubes over 4.
Methods: Make sure you are using the correct airway adapter for the size of trach tube. Make sure to zero calibrate new airway adapters
Troubleshooting:
- You will want to perform a zero calibration every time a new airway adapter is placed, when switching between sizes, if you receive the no breaths message or readings seem off. This must be done off the patient and input exposed to 0 mmHg CO2 concentration. To do this, click on the CO2, then Zero. The monitor will ask if it is exposed to 0 mmHg, click yes. You will see “zero in progress” on the screen then when the zero is finished, the numbers will flash and you are ready to use. If you receive a “zero error” message, repeat the process or try a new airway adapter.
- If you receive a “check adapter” message, replace airway adapter and do a zero calibration. Airway adapters are disposable items and need replaced approximately every 2-3 months. If an airway adapter is cracked or has been damaged, please replace. We recommend having 1-2 extra for when this happens.
- Alarm limits can be changed by clicking the blue CO2 tab
Last Modified: 04/22/2019