Troubleshooting:
Perform a zero calibration when placing a new sample line, when switching between sizes, when the readings seem off or you see a no breaths message.
This must be done off the patient and input exposed to 0 mmHg CO2 concentration. To do this, click on the CO2, then Zero. The monitor will ask if input is exposed to 0 mmHg, click yes. You will see “zero in progress” on the screen then when the zero is finished, the numbers will flash and you are ready to use. If you receive a “zero error” message, repeat the process or try a new sample line
If you receive an “occlusion” message, this means that moisture has hit the filter and you will need to replace sample line.
Sample lines are disposable items and need replaced approximately every 2-3 months. We recommend having 1-2 extra for when this happens.
When opening a new line, we recommend that you use a piece of tape and write the date on it.
Make sure T adapter is to the ceiling
Make sure monitor is above patient
Never tape a sample line together
Placement: There are high humidity lines which are used for general surgeries and low humidity for use in dentals. There are two sizes, pediatric (purple) for ET 4 and under and adult (clear) for ET sizes over 4.
Methods: Make sure you are using the correct sample line for type of surgery and trach tube size. Make sure to zero calibrate new sample lines, change lines in between patients, hang them with the mouth piece down to allow them time to dry out.
If occlusion message is seen, disconnect the sample line and replace with a new one. If mucous is seen in the line, dispose of line. If not, you can hang the sample line with the mouth piece down to allow time to dry and try again. If you still receive an occlusion message, throw the line away.
Care of Accessories: On average, the lines last around 2-3 months, unless you are a busy practice. Rotate sample lines in between patients and hang them with the mouth piece down to allow them time to dry out. If a sample line has visible mucous built up, please dispose of that line. If you receive an occlusion message and don’t see any visible moisture in the line, hang the line for 24 hours to dry out and try again. If the line is still occluded, please dispose of. Replace sample line if broken or cut
Cleaning: Never submerge sample lines in water or cleanser, the airway adapter can be cleaned with a kim wipe and a disinfectant.
Best Practices:
Attachments: Remove the sample line when finished and hang with mouth piece down to dry. This is the only accessory that should be removed daily.
Placement: There are high humidity lines which are used for general surgeries and low humidity for use in dentals. There are two sizes, pediatric (purple) for ET 4 and under and adult (clear) for ET sizes over 4.
Methods: Make sure you are using the correct sample line for type of surgery and trach tube size. Make sure to zero calibrate new sample lines, change lines in between patients, hang them with the mouth piece down to allow them time to dry out