LW8/LW9 has a white screen
Customer Service 561-689-0408

LW8/LW9 has a white screen

Troubleshooting:

1. The most common reason for this is that the monitor is not plugged in. Connect the monitor to power supply, wait for loading screen to come up. If it does not come up after 20 seconds, you will need to hold the toggle switch in to turn the monitor off or turn off the general power in the back of the monitor. 

2. If the battery no longer holds a charge, contact Digicare for service at 561-689-0408 Ext. 18.

Best Practices:

The monitor should always be plugged in when in use, the battery is meant as backup in the event of power failure or if the patient needs to be moved/transported.

Drain the battery once a month. To do this, unplug the monitor from the wall, turn the monitor on and let the battery drain out until the monitor shuts off. Connect to power supply to recharge when finished. 

Last Modified: 04/09/2019

Was this article helpful?

(1-Lowest, 10-Highest)